The LeSiever®️ LFT-1010 Integrity Tester is designed to meet the specific needs of filtration processes, backed by 25 years of integrity testing experience and developed by a domestic team.
This tester ensures reliable and comprehensive testing of your filters and filtration systems, confirming compliance with regulatory standards. The LeSiever®️ LFT-1010 integrates seamlessly into your processes, offering high reliability, ease of use, and automation to streamline operations. It adapts to various process requirements with flexible customization and adjustable settings, providing an efficient and intuitive testing solution.
Complete testing methods: capable of automatic bubble point,enhanced bubble point, and diffusion Flow, pressure attenuation,and water intrusion testing.
High testing accuracy and reproducibility: from small-scale filter membranes to large-scale filtration
Comprehensive power on self-test function: can check the air source pressure, internal electronic proportional valve, internal valve,internal sensor, airtightness, and electrical circuit for normal operation
Shorten testing time: Professional control circuit design and precise algorithm software reduce testing time by 20% to 30%.
Software independent upgrade: Customers can upgrade on site independently, eliminating the hassle of returning to the factory for upgrade and downtime, and reducing downtime.
The file has multiple retrieval functions; PDF files can be printed and exported.
A 15.6-inch large flat true color touch screen with comfortable viewing angle and IP54 protection level.
Scientific and convenient data management: USB drives can easily import and export PDF versions of data between testers and computers.
Fourth level user access permission management, making database usage and management more secure, organized, and in line with GMP production requirements.
It has a second level electronic signature, which has handwriting function and fully complies with GMP requirements.It also meets the requirements of 2FR Part 11 for electronic data integrity, electronic signature, and audit tracking.
Comprehensive data file management, including seven types of record files including “audit trail” and “instrument log”.
Through CE certification, the product complies with EU regulations.
Standard accessories
Air intake pipe (with quick connector), air outlet pipe (with quick connector and external exhaust valve), adapter, printing paper,power cord, operation manual.
Purchasing accessories
Special filter shell for filter element testing and flat filter for filter membrane testing.
Technical parameter
Test method
Bubble point, diffusion flow, water intrusion, pressure holding method, enhanced bubble point (diffusion flow+bubble point)
Pressure Unit
MPa,Psi,mbar
System resolution
0.0001 MPa
Test scope
Automatic bubble point 0.02 MPa -0.7 MPa
Pressure holding method 0-0.7 MPa
Diffusion flow 0.0-6500 mL/min
Water intrusion 0-200 mL/min
Testing accuracy
Bubble point: ≤ 0.005MPa (50 mbar)
Diffusion flow: ± 4% or ± 0.1ml/min, whichever is larger
Water intrusion: ± 4% or ± 0.02ml/min, whichever is larger
Pressure attenuation: ≤ 2%
Display
15.6 inch color touch screen
Input method Chinese and English input
Input method Chinese/English/Russian/Korean input
Printer
Equipped with a built-in thermal wide line printer or Thermal printer and dot-matrix printer, it can simultaneously or selectively print test parameters, results, and curves.
Storage Function
Ultra large capacity (1TB) memory, capable of supporting various file storage and backup requirements
Data output
USB drive or networking; Data formats: source data and PDF files
User management
Level 4 user permission management, password management
Ethernet transmission
Supporting networked transmission of MES, SCADA and other systems
Other functions
Data audit tracking function, automatically saving test data, hyperbolic display, monitoring the entire testing process, helping to analyze filter membranes
The performance of the filter element and filtration system, automatic self check and zero calibration, and on-site calibration.
Testing gas
Clean and dry compressed air or nitrogen
Source
AC100-240V,50/60Hz
Working temperature
5℃—40℃
Maximum intake pressure
0.8 MPa
External dimensions (mm)
377*270*297(长*宽*高)
Power dissipation
<200W
Working humidity
< 95% RH, No condensate water
Working pressure
0-0.7MPa(100PSI)
Net weight
10.5 kg
Ordering information
Product series
Article number
describe
Specification and model
LFT series
LFT-1010
HD touch widescreen with built-in operating system